Inter-Observer Reproducibility of Axial Ocular Measurements with Non-contact HAAG-STRAIT Biometer
Doi: 10.36351/pjo.v37i3.1239
DOI:
https://doi.org/10.36351/pjo.v37i3.1239Keywords:
Biometry, Axial Length, Central Corneal Thickness, Anterior Chamber Depth.Abstract
Purpose: To check inter observer reproducibility of axial ocular measurements i.e. central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), anterior segment lens (ASL), vitreous length (VL) and axial length (AL) with non-contact HAAG-STRAIT biometer.
Study Design: Comparative Reproducibility Analysis.
Place and Duration of Study: College of ophthalmology and allied vision sciences (COAVS), Mayo Hospital, Lahore.
Methods: This study included 66 healthy students (132 eyes). Data was collected through self-designed proforma by 2 operators independently. SPSS 21 was used for data analysis. Interclass correlation was applied for agreement between the two readings. Interclass coefficient (ICC) value greater than 0.7 was considered as excellent correlation.
Results: The mean CCT, AD, ACD, LT, ASL, VL, and AL were 526.47 ± 35.72 µm and 526.47 ± 36.06 µm
(ICC = 0.92); 2.93 ± 0.29 mm and 2.93 ± 0.29 mm (ICC = 0.81); 3.45 ± 0.30 mm and 3.46 ± 0.30 mm (ICC = 0.79); 3.58 ± 0.28 mm and 3.56 ± 0.22 mm (ICC = 0.76); 7.03 ± 0.30 mm and 7.02 ± 0.27 mm (ICC = 0.80); 16.56 ± 0.85 mm and 16.62 ± 0.81 mm (ICC = 0.72); and 23.59 ± 0.85 mm and 23.64 ± 0.87 mm (ICC: 0.76) of observer 1 and 2, respectively.
Conclusion: Non-contact Biometer (HAAG-STRAIT) has high inter-observer reproducibility with strong interclass coefficient of greater than 0.72.
Key Words: Biometry, Axial Length, Central Corneal Thickness, Anterior Chamber Depth.